Gamma camera

Gamma camera

A gamma camera is a device used to image gamma radiation emitting radioisotopes, a technique known as scintigraphy. The applications of scintigraphy include early drug development and nuclear medical imaging to view and analyse images of the human body of the distribution of medically injected, inhaled, or ingested radionuclides emitting gamma rays.

Construction

A gamma camera consists of one or more flat crystal planes or, detectors, optically coupled to an array of photomultiplier tubes, the assembly is known as a "head", mounted on a gantry. The gantry is connected to a computer system that both controls the operation of the camera as well as acquisition and storage of acquired images.

The system accumulates events, or counts, of gamma photons that are absorbed by the crystal in the camera. Usually a large flat crystal of sodium iodide with thallium doping in a light-sealed housing is used. The highly efficient capture method of this combination for detecting gamma rays was discovered by noted physicist Robert Hofstadter in 1948 [http://nobelprize.org/nobel_prizes/physics/laureates/1961/hofstadter-bio.html] ).

The crystal scintillates in response to incident gamma radiation. When a gamma photon leaves the patient (who has been injected with a radioactive pharmaceutical), it knocks an electron loose from an iodine atom in the crystal, and a faint flash of light is produced when the dislocated electron again finds a minimal energy state. The initial phenomenon of the excited electron is similar to the photoelectric effect and (particularly with gamma rays) the Compton effect. After the flash of light is produced, it is detected. Photomultiplier tubes (PMTs) behind the crystal detect the fluorescent flashes (events) and a computer sums the counts. The computer reconstructs and displays a two dimensional image of the relative spatial count density on a monitor. This reconstructed image reflects the distribution and relative concentration of radioactive tracer elements present in the organs and tissues imaged.

ignal processing

Hal Anger developed the first gamma camera in 1957. His original design, frequently called the Anger camera, is still widely used today. The Anger camera uses sets of vacuum tube photomultipliers. Generally each tube has an exposed face of about 3 inches in diameter and the tubes are arranged in hexagon configurations, behind the absorbing crystal. The electronic circuit connecting the photodetectors is wired so as to reflect the relative coincidence of light fluorescence as sensed by the members of the hexagon detector array. All the PMTs simultaneously detect the (presumed) same flash of light to varying degrees, depending on their position from the actual individual event. Thus the spatial location of each single flash of fluorescence is reflected as a pattern of voltages within the interconnecting circuit array.

The location of the interaction between the gamma ray and the crystal can be determined by processing the voltage signals from the photomultipliers; in simple terms, the location can be found by weighting the position of each photomultiplier tube by the strength of its signal, and then calculating a mean position from the weighted positions. The total sum of the voltages from each photomultiplier is proportional to the energy of the gamma ray interaction, thus allowing discrimination between different isotopes or between scattered and direct photons.

patial resolution

In order to obtain spatial information about the gamma emissions from an imaging subject (e.g. a person's heart muscle cells which have absorbed an intravenous injected radioactive, usually thallium-201 or technetium-99m, medicinal imaging agent) a method of correlating the detected photons with their point of origin is required.

The conventional method is to place a collimator over the detection crystal/PMT array. The collimator consists of a thick sheet of lead, typically 1-3 inches thick, with thousands of adjacent holes through it. The individual holes limit photons which can be detected by the crystal to a cone; the point of the cone is at the midline center of any given hole and extends from the collimator surface outward. However, the collimator is also one of the sources of blurring within the image; lead does not totally attenuate incident gamma photons, there can be some crosstalk between holes.

Unlike a lens, as used in visible light cameras, the collimator attenuates most (>99%) of incident photons and thus greatly limits the sensitivity of the camera system. Large amounts of radiation must be present so as to provide enough exposure for the camera system to detect sufficient scintillation dots to form a picture.

Other methods of image localization (pinhole, rotating slat collimator with CZT (Gagnon & Matthews) and others) have been proposed and tested; however, none have entered widespread routine clinical use.

The best current camera system designs can differentiate two separate point sources of gamma photons located a minimum of 1.8 cm apart, at 5 cm away from the camera face. Spatial resolution decreases rapidly at increasing distances from the camera face. This limits the spatial accuracy of the computer

canning and use in PET

SPECT (single photon emission computed tomograpy) imaging, as used in nuclear cardiac stress testing, is performed using gamma cameras, usually one, two or three detectors or heads, are slowly rotated around the patient's torso.

Multi-headed gamma cameras can also be used for Positron emission tomography scanning, provided that their hardware and software can be configured to detect 'coincidences' (near simultaneous events on 2 different heads). Gamma camera PET is markedly inferior to PET imaging with a purpose designed PET scanner, as the scintillator crystal has poor sensitivity for the high-energy annihilation photons, and the detector area is significantly smaller. However, given the low cost of a gamma camera and its additional flexibility compared to a dedicated PET scanner, this technique is useful where the expense and resource implications of a PET scanner cannot be justified.

ee also

*Nuclear medicine

References

* H. Anger. A new instrument for mapping gamma-ray emitters. Biology and Medicine Quarterly Report UCRL, 1957, 3653: 38. (University of California Radiation Laboratory, Berkeley)
* Anger HO. Scintillation camera with multichannel collimators. J Nucl Med 1964 Jul;65:515-31. PMID 14216630
* PF Sharp, et al, Practical Nuclear Medicine, IRL Press, oxford
* D. Gagnon, C.G. Matthews, US Patent #6,359,279 & #6,552,349
* Physics in nuclear medicine, third edition, Cherry, Sorenson, Phelps

"'


Wikimedia Foundation. 2010.

Игры ⚽ Поможем написать реферат

Look at other dictionaries:

  • gamma camera — n a camera that detects the gamma ray photons produced by radionuclide decay and is used esp. in medical diagnostic scanning to create a visible record of a radioactive substance injected into the body * * * a piece of apparatus that detects… …   Medical dictionary

  • Gamma caméra — Scintigraphie La scintigraphie est une méthode d imagerie médicale qui procède par l administration, dans l organisme, d isotopes radioactifs afin de produire une image médicale par la détection des rayonnements émis par ces isotopes après… …   Wikipédia en Français

  • gamma camera — noun : a camera that detects the gamma ray photons produced by radionuclide decay and is used especially in medical diagnostic scanning to create a visible record of a radioactive substance injected into the body * * * gamma camera noun A device… …   Useful english dictionary

  • gamma camera — /gæmə ˈkæmrə/ (say gamuh kamruh) noun a camera which is sensitive to gamma rays emitted by radionuclides, used in nuclear medicine …  

  • gamma camera — a piece of apparatus that detects radioactivity in the form of gamma rays emitted by radioactive isotopes that have been introduced into the body as tracers. It contains an activated sodium iodide crystal (see scintillator) and a large array of… …   The new mediacal dictionary

  • gamma camera — noun Date: 1964 a camera that detects the radiation from a radioactive tracer injected into the body and is used especially in medical diagnostic scanning …   New Collegiate Dictionary

  • Gamma ray — Gamma rays (denoted as gamma;) are a form of electromagnetic radiation or light emission of frequencies produced by sub atomic particle interactions, such as electron positron annihilation or radioactive decay. Gamma rays are generally… …   Wikipedia

  • Caméra à scintillation — Scintigraphie La scintigraphie est une méthode d imagerie médicale qui procède par l administration, dans l organisme, d isotopes radioactifs afin de produire une image médicale par la détection des rayonnements émis par ces isotopes après… …   Wikipédia en Français

  • Gamma correction — Gamma correction, gamma nonlinearity, gamma encoding, or often simply gamma, is the name of a nonlinear operation used to code and decode luminance or tristimulus values in video or still image systems. [cite book | title = Digital Video and HDTV …   Wikipedia

  • Gamma ray burst — [ GRB 990123 taken on January 23, 1999. The burst is seen as a bright dot denoted by a square on the left, with an enlarged cutout on the right. The object above it with the finger like filaments is the originating galaxy. This galaxy seems to be …   Wikipedia

Share the article and excerpts

Direct link
Do a right-click on the link above
and select “Copy Link”